High-Performance Multifunctional Microplate Reader BET96FAL - China Suppliers & Factory for Absorbance, Fluorescence, Luminescence
Basic Parameters and Performance
Model: ELx808IUBET
Absorbance Specifications
Light Source
Tungsten halogen
Wavelength Selection
Filters
Wavelength Range
405/450/490/630/340 nm (ELx808IU)
Bandpass
10 nm
Dynamic Range
0 – 4.0 OD
Resolution
0.001 OD
Filter Wheel Capacity
6 positions
Filters Supplied
5 filters (5 with UV option)
OD Accuracy
<1% at 2.5 OD
<2% at 3.5 OD
<2% at 3.5 OD
OD Linearity
<1% at 2.5 OD
OD Repeatability
<0.5% at 2.5 OD
<1.5% at 3.5 OD
<1.5% at 3.5 OD
Reading Speed
96 wells: 8 seconds
Physical Characteristics
Power Supply
100 – 240 Volts AC, 50/60 Hz
Dimensions
16”D x 15.5”W x 8.75”H
(40.6 x 39.4 x 22.2 cm)
(40.6 x 39.4 x 22.2 cm)
Weight
30 lbs (13.6 kg)
Frequently Asked Questions
What is the measurement speed of the ELx808IUBET?
The ELx808IUBET delivers rapid throughput, capable of reading a complete 96-well microplate in just 8 seconds.
What wavelength channels and filters are available?
The unit features a 6-position filter wheel supplied with 5 standard filters including the UV option: 405 nm, 450 nm, 490 nm, 630 nm, and 340 nm, with a 10 nm bandpass.
What are the dynamic range and resolution performance?
It provides a dynamic measurement range of 0 to 4.0 OD with high-precision optical resolution down to 0.001 OD.
How accurate and repeatable are the optical density (OD) measurements?
OD accuracy is <1% at 2.5 OD and <2% at 3.5 OD. OD linearity is <1% at 2.5 OD, and OD repeatability is <0.5% at 2.5 OD and <1.5% at 3.5 OD.
What are the electrical power specifications?
The instrument features universal voltage compatibility supporting 100 – 240 Volts AC at 50/60 Hz.
What are the overall dimensions and weight of the device?
The unit measures 16”D x 15.5”W x 8.75”H (40.6 x 39.4 x 22.2 cm) and weighs approximately 30 lbs (13.6 kg).