BETPro96X Endotoxin Reader (Absorbance Kinetic Incubating Microplate Reader)
Instrument features
1) Adopt 7-inch high-resolution LCD touch screen, no keyboard required, simple and intuitive operation, easy to use.
2) 8-bit filter wheel, 4 filters as standard, optional other filters; wavelength range is 340-1000nm, which can meet a variety of applications;
3) 9-channel vertical light path, automatic positioning function of plate hole center.
4) Single wavelength, dual wavelength detection, fast plate reading; it only takes 5 seconds to read all 96-well plates or 11 seconds to complete dual wavelength detection.
5) The touch screen software with modular design is convenient for endpoint method and kinetic analysis, and different detection steps, vibration plate steps, temperature control steps and program comments can be added multiple times. Support USB export data and direct printing with thermal printer.
6) It has self-test and diagnosis functions for light path, moving position, etc.
7) It has vibration plate function, and the vibration plate time and speed are adjustable.
8) Energy-saving design of light source to maximize the service life of light source.
9) Multiple data output interfaces for convenient data transmission and processing.
10) Unique communication design, can be directly connected to common Lis systems or analysis and control software on the market, such as: Ruimei, Lianqiao and other list systems, Idexx Chek, RIDAWIN, PriboFast, EndoScanV, Win QCL, Gen5, KC4, KCJunior and other analysis and control software.
Basic parameters and performance
Model: BETPro96X
Light source: 6V10W halogen lamp (lifespan: 6000 hours)
Wavelength range: 340~1000nm
Filter: standard configuration 340, 405, 450, 490nm filters, up to eight filters can be installed
Absorbance range: 0.000-5.000Abs
Resolution: 0.001Abs
Linearity: (0, 2.0) ≤±1%; [2.0, 5.0] ≤±2%
Wavelength indication error: ≤±3nm
Wavelength repeatability: ≤±1.5nm
Repeatability: [0, 3.0] ≤0.3%; [3.0, 5.0] ≤1%
Stability: [0, 3.0] ≤0.3%; [3.0, 5.0] ≤1%
Accuracy: ≤±1.0% ±0.010 OD from 0.000 to 2.500 OD @405nm
Sensitivity: Sensitivity ≥ 0.010
Channel difference: <0.01A (1.0Abs standard filter)
Measurement speed: single wavelength <10s/96 wells, dual wavelength <16s/96 wells (normal mode)
Incubation temperature: room temperature + 4℃ to 65℃, with hot cover to prevent agglutination technology
Shaking: Liner shaking with low, medium, high and variable speeds.
Instrument Size: 278x430x218mm(WxDxH)
Instrument Net Weight: 9.8kg